The aim of implanting a leadless cardiac pacemaker is to detect cardiac bradyarrhythmias and deliver electric pulses to the heart to increase the heart rate. The leadless pacemaker has a built‑in pulse generator, battery and electrodes. The procedure is done under local anaesthesia, with or without sedation, in a cardiac catheterisation laboratory. Under fluoroscopic guidance, the proximal end of the pacemaker is attached to a deflectable bespoke delivery catheter system and inserted percutaneously through the femoral vein using a dedicated introducer sheath. It is then advanced into the right atrium through the tricuspid valve, into the right ventricle and positioned near the apex or lower septum. Contrast may be injected into the right ventricle to visualise the desired location. Once positioned, the pacemaker is deployed and securely implanted into the endocardial wall using a fixation mechanism (a screw‑in helix or nitinol tines). An electrode at the distal end of the pacemaker delivers electrical impulses that pace the heart. Electrical measurements are taken and, if satisfactory, the pacemaker is released from the catheter and the catheter is removed. If the position is suboptimal, the pacemaker can be detached from the endocardium and repositioned prior to final release of the delivery catheter.